Determination of 16 Harmful Substances in Solvent Based Coatings by GC-MS Method
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:solvent based coatings  harmful substances  GC-MS  sonicate extraction
  
AuthorInstitution
薛希妹,薛秋红,刘心同,单宝田,牛增元 1.中国海洋大学化学化工学院;2.山东出入境检验检疫局
Hits: 2666
Download times: 1034
Abstract:
      A GC-MS method was developed for the determination of 16 kinds of harmful substances(eg.methanol,volatile halohydrocarbons,benzene homologues,the free toluene diisocyanate and hexane diisocyanate)in solvent based coatings.The extraction and separation efficiencies of targeted compounds by using organic solutions of ethyl acetate,n hexane,tetrahydrofuran and acetonitrile were investigated.The procedure of sample pre treatment and the conditions of instruments were optimized.The quantitative analysis of compounds were carried out by using 2-bromopropane and 1,2,4-trichlorobenzene as the internal standard,and acetonitrile as extraction reagent.The result indicated that the calibration curves of the 16 harmful substances were linear in the range of 5-200 mg/L with correlation coefficients no less than 0.999.The spiked recoveries ranged from 80% to 105% with relative standard deviations less than 5.0%.The detection limits ranged from 0.08 mg/L to 1.41 mg/L.The method was rapid,sensitive and accurate.