Direct Electrochemistry of Myoglobin on Nafion/Fe3O4-CdTe/CdS Composite Film and Its Application in Hydrogen Peroxide Biosensing
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:myoglobin  Fe3O4  CdTe/CdS QDs  hydrogen peroxide  biosensor
  
AuthorInstitution
赵倩,王国涛,贾丽萍,陶绪泉,王怀生 1.聊城大学化学化工学院,2.聊城大学材料科学与工程学院
Hits: 2279
Download times: 1036
Abstract:
      A Nafion/Fe3O4-CdTe/CdS-myoglobin(Mb)modified glassy carbon electrode(GCE) was prepared by immobilizing Mb on the electrode surface using Fe3O4 nanoparticles and CdTe/CdS quantum dots combining the advantages of film formation ability of Nafion.Cyclic voltammetric data showed that the immobilized Mb exerted a good direct electrochemistric behavior and a good stability with a pair of well defined and nearly reversible peaks at about-0.351 V in 0.1 mol/L pH 7.0 PBS corresponding to the characteristic peaks of the Mb heme Fe(Ⅲ)/Fe(Ⅱ) redox couples.The results indicated that the microenvironment of Nafion/Fe3O4-CdTe/CdS film could promote the direct electron transfer between Mb heme Fe(Ⅲ)/Fe(Ⅱ) and GCE and provide the matrix for the immobilization of Mb.The electrocatalytic ability of the immobilized Mb to the reduction of H2O2 was also investigated and the results suggested that the modified electrode could be used as a biosensor for the detection of hydrogen peroxide with high sensitivity(30.6 mA·L/mol),low detection limit( 0.89 μmol/L,S/N=3) and rapid response.
Close