Determination of 8 Phthalate Esters in Liquid Cosmetics Using Solid-phase Microextraction/Gas Chromatography-Mass Spectrometry
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:Solid-phase microextraction  gas chromatography-mass spectrometry  liquid cosmetics  phthalic acid esters  plasticizer
  
AuthorInstitution
张璇,陈大舟,汤桦,冯洁,庄惠明,李蕾 1.北京化工大学理学院化学系; 2.中国计量科学研究院化学计量与分析科学研究所;3.云南省计量测试技术研究院
Hits: 2616
Download times: 1038
Abstract:
      A method based on solid-phase microextraction(SPME) technique followed by gas chromatography with mass spectrometry detector(GC-MS) was developed for the determination of 8 phthalic acid esters(PAEs) in liquid cosmetics.The parameters affecting the partitions of analytes such as fiber coating,extraction temperature,stirring speed and salt concentration were optimized.The final SPME method employed a 65 μm polydimethylsiloxane(PDMS)/divinylbenzene(DVB) coating fiber and the saturated sodium chloride solution at 90 ℃ under stir for 60 min,then the fiber was introduced into the GC injector port and the extract was desorbed at 250 ℃ for 4 min.The detection limits for the other 7 PAEs except for DPhP ranged from 0.7 μg?kg-1 to 13.6 μg?kg-1,and the recoveries for spiked samples were between 83% and 97% with RSDs(n=5)of 2.5%-10.0% .For its weak adsorption to fiber coating,the detection limit for DPhP was 75 μg?kg-1 ,and the recovery was 70% with RSD(n=5) of 13.9%.The method was applied in the determination of 8 PAEs in liquid samples with satisfactory results.