Control System Design and Application of Fast Gas Chromatography
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:fast gas chromatography  temperature programming  control system
  
AuthorInstitution
孙宁,谢春光,麦泽彬,吴曼曼,张业荣,谭国斌,郭长娟,黄正旭,高伟 1.上海大学环境与化学工程学院环境污染与健康研究所;2.广州禾信分析仪器有限公司;3.华南师范大学化学与环境学院;4.暨南大学大气环境安全与污染控制工程研究所
Hits: 2993
Download times: 1124
Abstract:
      A new control system was designed for fast gas chromatography(FGC) in this paper.The controller is mainly composed of the column temperature control system,automatic sampling and pressure control system of the gas path.The range of temperature control is from 30 to 160 ℃,the heating rate approximates 3 ℃/s,the temperature control accuracy is about ± 0.5 ℃,and the carrier gas pressure control range is from 0 to 0.5 MPa.Applying the controller in homemade FGC,a standard sample composed of straight-chain n-alkanes(C1-C8) and toluene was tested by the FGC.The result indicated that the 9 species could be completely separated in 100 s.