Rapid Determination of Glucosamine Content by High-performance Liquid Chromatography
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:glucosamine sulfate potassium chloride capsules  glucosamine  high-performance liquid chromatography(HPLC)  determination
  
AuthorInstitution
谢志刚,李博,陈书军,彭秀兰,王莹,尹鸿萍,奚涛 1.中国药科大学生命科学与技术学院;2.江苏双林海洋生物药业有限公司
Hits: 3904
Download times: 2194
Abstract:
      A high-performance liquid chromatography(HPLC) method for the determination of glucosamine in glucosamine sulfate potassium chloride capsules was developed.The separation of analyte was performed on a Hedera NH2 column (250 mm ×4.6 mm,5 μm) with acetonitrile and phosphate buffer (65∶35) as mobile phase,at a flow rate of 1.0 mL/min.The variable wavelength detector(VWD) was set at 195 nm.In the above chromatographic conditions,the calibration curve was linear over glucosamine concentration in the range of 1.6-2.4 g/L.The limit of detection(LOD) of glucosamine was 7.81 mg/L. The RSD(n=6) for the accuracy was 0.53% and the average recovery was 100%with RSD of 1.1%.This method is simple,accurate and with no need of derivation,and is suitable for the determination of glucosamine.