Preparation of Silica-coated Magnetic Nanomaterials Modified with Ionic Liquid and Its Application in Detection of Trace Cadmium Ion in Water
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:magnetic nanomaterials  ionic liquid  cadmium ion  water sample
  
AuthorInstitution
党永锋,马晓国,邹建平 广东工业大学环境科学与工程学院
Hits: 2638
Download times: 1122
Abstract:
      A kind of silica-coated magnetic nanomaterials modified with 1-octyl-3-methylimidazolium hexafluorophosphate([OMIM]PF6) was prepared,and characterized by scanning electron microscope(SEM),Fourier transform infrared spectroscopy(FT-IR),vibrating sample magnetometer(VSM) and thermogravimetric analyzer(TGA).The synthetic material(Fe3O4-SiO2-IL) showed a good acid resistance.The adsorption behavior of cadmium ion onto Fe3O4-SiO2-IL was tested,and it was found that the equilibrium adsorption followed Langmuir isotherms well.Based on this,a new method for separation and preconcentration of Cd2+ in aqueous solution was developed by using modified silica-coated magnetic nanoparticle as the sorbent and flame atomic absorption spectrometry(FAAS) as detector.Effects of some factors on recovery of Cd2+ were investigated.The optimum conditions were as the follows:pH 6.0;eluent:1 mol/L HCl;oscillation time:10 min;standing time:20 min.Under the optimized conditions,the limit of detection(LOD) for Cd2+ was 0.2 μg/L,and relative standard deviation(RSD,n=6) for 15 μg/L Cd2+ was 2.5%.The proposed method was applied in the determination of Cd2+ in river and lake water with spiked recoveries of 84%-100%.