Selective and Sensitive Determination of Uric Acid on PABSA/GN Composite-modified Glassy Carbon Electrode
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:p-aminobenzene sulfonic acid  graphene nanosheets  modified electrode  selective determination  uric acid(UA)  ascorbic acid(AA)
  
AuthorInstitution
孙娜,王宗花,夏建飞,郭新美,张菲菲,夏延致,李延辉 1.青岛大学纤维新材料与现代纺织实验室国家重点实验室培育基地;2.青岛大学化学化工与环境学院
Hits: 2436
Download times: 979
Abstract:
      A PABSA/GN/GCE composite-modified electrode was fabricated by using a graphene nanosheets(GN) modified glassy carbon electrode(GCE)electropolymerized with p-aminobenzene sulfonic acid(PABSA).The electrochemical behaviors of uric acid(UA) and ascorbic acid(AA) on the nanocomposites-modified electrode were investigated.Compared with PABSA/GCE and GN/GCE,the PABSA/GN/GCE improved the resolution and the sensitivity for the determination of UA in the presence of AA.In 0.1 mol/L phosphate buffer solution(pH 7.0),a peak to peak separation of 344 mV was observed,which enabled the selective determination of UA.The peak currents obtained from DPV were linearly related to concentrations of UA in the range of 1.0×10-7-8.0×10-4 mol/L.The detection limit was 4.5×10-8 mol/L.The nanocomposites-modified electrode was adopted for the detection of UA in the human urine with satisfactory result.