Study on Flow Injection Spectrophotometry Determination of Free Chlorine in Running Water
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:free chlorine  flow injection  DPD spectrophotometry  on-line monitor
  
AuthorInstitution
孙西艳,付龙文,周娜,冯巍巍,吕颖,陈令新 中国科学院烟台海岸带研究所海岸带环境过程重点实验室
Hits: 2618
Download times: 1195
Abstract:
      A flow injection on-line analyzer for the determination of free chlorine was designed,and the optimal condition was investigated.Based on the improved N,N-diethyl-1,4-phenylenediamine(DPD) colorimetic principle,the analyzer used flow injection sampling technology,combined with photoelectric conversion,digital signal processing and automatic control technology.It could overcome the defects of DPD spectrophotometry done by manual operation,and perform automated sample analysis.The method was proven to consume less reagent with wide linear range(0.05-6.00 mg/L) and lower detection limit(0.019 mg/L).The recoveries were in the range of 90%-98%.The analyzer was apt to be widely used in monitoring free chlorine in the waterworks and water supply networks with rapid sampling frequency(60 h-1) and higher precision(RSD≤1.2%).