Aptamer-based Chemiluminescent Detection of Adenosine
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:chemiluminescence  adenosine  aptamer  magnetic bead  horseradish peroxidase
  
AuthorInstitution
鲁卢,严喜鸾,肖义陂,陈静雯,肖鉴谋 1.南昌大学环境与化学工程学院;2.南昌市洪都中医院
Hits: 2862
Download times: 1691
Abstract:
      Based on magnetic beads as separating carrier and adenosine as a specific competitor appeared,a method of aptamer-based chemiluminescence(CL) for the determination of adenosine was developed.Experimental parameters including magnetic beads,adenosine-aptamer and report DNA were optimized.Under the optimal conditions,the decreased CL intensity was proportional to the adenosine concentration in the range of 1.0×10-9-1.0×10-4 mol/L(r2=0.99),with a detection limit of 1.0×10-9 mol/L.The relative standard deviation(n=5) for 0.1 mmol/L adenosine is 5.0%.The present method was successfully applied in the determination of adenosine in real human urine.The method is simple,sensitive and specific for the detection of adenosine,and it is a promising approach for the determination of small biomolecules in the fields of clinical diagnosis,pharmaceutical analysis and so on.
Close