Determination of Trace Benzenes in Water Samples by On-line Thermal Desorption-Gas Chromatography after Solid Phase Extraction
  
View Full Text    Download reader
DOI:
KeyWord:Fe/SiO2/PDMS  benzenes  gas chromatography  electromagnetic induction heating  solid phase extraction  on line thermal desorption
  
AuthorInstitution
LIU Bai-zhan,XIE Shuang-shuang,ZHENG Sai-jing,LI Gang,WANG Wei-miao,YANG Jun,GAN Wu-er* 1.上海烟草集团有限责任公司烟草行业卷烟烟气重点实验室;2.中国科学技术大学化学系
Hits: 2505
Download times: 1279
Abstract:
      A novel method was developed for the determination of trace benzenes in water samples by on-line thermal desorption-gas chromatography after solid phase extraction on Fe/SiO2/PDMS particles bed column.The solid phase extraction column was prepared by filling prepared Fe/SiO2/PDMS particles into quartz tube.The adsorbed benzenes on the column was thermally desorbed and directly introduced into gas chromatography for separating and determing.Several factors affecting the extraction and separation efficiencies were investigated and optimized. Under the optimal conditions,the calibration curves of benzenes were linear in the range of 0.1-20 ng/L,with corrrelation coefficients(r) of 0.999 2-0.999 6 and detection limits(S/N=3) of 0.03-0.05 ng/L.The relative standard deviations(RSD,n=11) for 1 ng/L benzenes were in the range of 2.0%-4.4%.The presented method was sensitive and accurate,and was successfully applied in the determination of benzenes in water samples.
Close