Spectrofluorimetric Determination of Indole Phytohormones in Bean Sprout Using Three Phase Hollow Fiber-Liquid Phase Microextraction
  
View Full Text    Download reader
DOI:
KeyWord:hollow fiber three phase liquid phase microextraction  indole phytohormones  orthogonal experiment  spectrofluorimetry  bean sprout
  
AuthorInstitution
CHEN Lu, TURGHUN Muhammad*, MUNIRA Abuduani, SHIRALI Jappar 新疆大学化学化工学院石油天然气重点实验室
Hits: 2329
Download times: 1269
Abstract:
      A new method using three phase hollow fiber-liquid phase microextraction was developed for the determination of indole phytohormones in bean sprout by spectrofluorimetry.The parameters affecting liquid phase microextraction performance were optimized using L9(34)orthogonal experiment.The optimum extraction was achieved by adjusting sample solution pH value to 4.0, using 1-octanol as organic extraction solvent, pH 12.0 solution as acceptor phase, and stirring the solution at 1 000 r/min for 60 min.Under the optimum conditions, the calibration curve showed a good linearity(r=0.997 9) in the range of 1.71-50.0 mg/L, with a detection limit of 0.57 mg/L.Recoveries for spiked samples from 88.6%to 100.7%were obtained with RSDs not more than 4.8%.The proposed method was simple, green and accurate in the determination of indole phytohormones in bean sprout.
Close