Determination of Aflatoxins in Feeds by High Performance Liquid Chromatography-Quadrupole/Electrostatic Field Orbitrap High Resolution Mass Spectrometry with Online Solid Phase Extraction
  
View Full Text    Download reader
DOI:
KeyWord:aflatoxins  online solid phase extraction  high performance liquid chromatography-quadrupole/electrostatic field Orbitrap mass spectrometry(HPLC-Q/Orbitrap MS)  feed
  
AuthorInstitution
LIU Fei,REN An-shu,GE Ping,DING Nian-fang,NAREN Qi-qi-ge 南京华谱分析检测技术服务有限公司
Hits: 2159
Download times: 875
Abstract:
      A high performance liquid chromatography-quadrupole/electrostatic field Orbitrap high resolution mass spectrometry(HPLC-Q/Orbitrap HRMS) was developed for the determination of aflatoxins in feeds.The IMMUNOPREP- ONLINE AFLATOXIN(Part Code:P900) was used as an online solid phase extraction column,and DiamonsilPlus C18(150 mm×4.6 mm,5 μm) as an analytical column.The feed samples were dissolved with a certain amount of NaCl and acetonitrile-water(80∶20).After ultrasonic extraction,the supernatant was injected into an online solid phase extraction column,then the target was transferred to the analysis column by valve switching technique and HPD confocal elution mode for separation.The external standard method was used for quantitation.The analysis was performed by MS under Full scanning mode in positive mode.Under the optimal conditions,the calibration curves were linear in the range of 0.5-50.0 μg/L for four aflatoxin,the limits of detection and quantitation were 0.2 μg/kg and 0.5 μg/kg,respectively.The average recoveries at three spiked levels(0.5,1.0,5.0 μg/kg) were in the range of 94.6%-114.3% with relative standard deviations(RSDs) not more than 8.3%.With the advantages of rapidness,high automation,high-throughput detection and low cost,the proposed method could be used in the rapid detection on aflatoxins in feed samples.
Close