Study on Adsorptive Voltammetric Detec? tion of Coenzyme Q10
  
View Full Text    Download reader
DOI:
KeyWord:coenzyme Q10  adsorptive cyclic voltammetry  response surface optimization  quantita? tive analysis
  
AuthorInstitution
LIU Yu-hong,GUO Bin,TU Yi-feng 1. College of Pharmacy,Medical University of Jinzhou,Jinzhou ,China;2. College of Chemistry, Chemical Engineering and Material Science,Soochow University,Suzhou ,China
Hits: 1132
Download times: 633
Abstract:
      The electrochemical behavior of coenzyme Q10(CoQ10) on glassy carbon electrode (GCE)was explored by cyclic voltammetry(CV)in this paper. The results revealed the adsorption of CoQ10 on GCE,thus providing an approach to enhance the sensitivity for its detection by pre-enrich? ment. The experimental conditions including ultrasonic duration for CoQ10 extraction,detection tem? perature and the ratio of electrolyte solvent(methanol-ethanol)were optimized by Box-Behnken method as 1 min 50 s,25 ℃ and 1. 1∶1,respectively. Under the optimized conditions,after 25 min of adsorption,the CV current linearly responded. to the logarithm of CoQ10 concentration within the range of 0. 025-25. 0 μmol/L,with a regression equation of I = 2. 554 + 1. 479lgC(r 2 = 0. 997),and a detection limit(LOD)of 19. 0 nmol/L. The detection of CoQ10 by using adsorptive CV on GCE could achieve its accurate and high response quantitative analysis.
Close